Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 43
Show Transmitter Status
Use the STATUS command to view detailed information on transmitter
model and configuration.
STATUS [function]<cr>
where
Function
=
Optional switch to display a more detailed status for
BACnet or Modbus protocol
(available on transmitter
models with digital output).
Available switches
are -bacnet and modbus.
Example (display detailed BACnet status):
>status bacnet
* BACnet module (BACNET) *
BACnet protocol : active
MAC : 0 (00h)
Device Instance : 6 (00000006h)
Name : NOT_SET
Location : Location
Description : Description
MAX_MASTER : 127 (7Fh)
Node type : Master
Baud setting : Auto
Current baudrate : 19200 8N1
Baudrate locked : No
Baud detection interval: 10 s
DCC : Communication enabled
Valid frames : 0
Invalid frames : 0
Unwanted frames : 0
Lost tokens : 0
Failed TX : 0