EasyManua.ls Logo

Vaisala HUMICAP HMW90 SERIES - Dimensions in mm

Vaisala HUMICAP HMW90 SERIES
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
84 ___________________________________________________________________ M211399EN-F
Table 16 Inputs and Outputs
Property
Description / Value
HMW92 and TMW92
Outputs
HMW92
2 x 4 ... 20 mA, loop powered
TMW92
1 x 4 ... 20 mA, loop powered
Loop resistance
0 ... 600 Ω
Supply voltage
20 ... 28 VDC at 500 Ω load
10 .... 28 VDC at 0 Ω load
Isolation between output channels
500 VDC
HMW93 and TMW93
Outputs
HMW93
2 x 0 ... 5V, 0 ... 10 V
TMW93
1 x 0 ... 5V, 0 ... 10 V
Load resistance
10 kΩ min.
Supply voltage
18 ... 35 VDC
24 VAC ±20 % 50/60 Hz
Max. current consumption
12 mA
max. with relay 25 mA
Relay
1 pc (SPST, max. 50 VDC, 500 mA)
3-wire installation max cable
resistance
2.5 Ω at 24V supply
(with 10 V output, relay not used)
HMW95
Supply voltage
18 ... 35 VDC
24 VAC ±20 % 50/60 Hz
Current consumption (with
termination)
Average
10 mA at 24 VDC
Maximum
30 mA at 24 VDC
Power consumption
< 0.3 W
Output type
RS-485 (galvanic isolation, 1.5 kV)
RS-485 end of line termination
Enable with jumper, 120 Ω
Supported protocols
Selectable by DIP switch
BACnet MS/TP
Operating mode
Selectable Master/Slave
Address range, master mode
0 ... 127
Address range, slave mode
128 ... 255
Modbus RTU
Address range
0 ... 247
Service port
RS-485 line
for temporary service use
Table 17 Mechanics
Property
Description / Value
IP class
IP30
Standard housing color
White (RAL9003*)
Optional housing colors
(configurable models only)
Black (RAL9005*)
Grey (RAL7035*)
Light Ivory (RAL1015*)
Housing material
ABS/PC, UL-V0 approved
Output connector
Screw terminals
max. wire size 2 mm
2
(AWG14)
Service port connector
4-pin M8
Weight
155 g

Table of Contents

Related product manuals