EasyManua.ls Logo

Vaisala HUMICAP HMW90 SERIES - Device Identification Objects; Exception Responses

Vaisala HUMICAP HMW90 SERIES
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B _________________________________________________________ Modbus Reference
VAISALA _______________________________________________________________________ 105
NOTE
A complete 32
-bit floating point value should be read and written in a
single
Modbus transaction (for example, function codes 05, 06
, and 22 do
not have an effect on floating point values).
16-Bit Integer Format
16-bit integer values are scaled to include the necessary decimals (see
corresponding register table for the scaling factor) and represented as
described in the following table.
Table 41 16-bit signed integer format details
Value
Description
0x0000…0x7FFE
Value in range 0…32766
0x8002…0xFFFF
Value in range -32766…-1 (2’s complement)
0x7FFF
Value is 32767 or larger (positive infinity)
0x8000
Value is not available (quiet NaN)
0x8001
Value is -32767 or smaller (negative infinity)
Trying to write special values 0x7FFF, 0x8000, or 0x8001 in a 16-bit
integer register is ignored to avoid unintended result.
NOTE
Some values may exceed the signed
16-bit range even in normal
operation. Please access such values using the floating point registers
instead.
Measurement Data (Read-Only)
Table 42 HMW90 Modbus Measurement Data Registers
Name
Metric float
Metric integer
Metric
unit
Non-metric
float
Non-metric integer
Non-metric
unit
RH
0003…0004
0258 (×0.01)
%RH
6403…0004
6458 (×0.01)
%RH
T
0005…0006
0259 (×0.01)
°C
6405…0006
6459 (×0.01)
°F
Td
0007…0008
0260 (×0.01)
°C
6407…0008
6460 (×0.01)
°F
Tdf
0009…0010
0261 (×0.01)
°C
6409…0010
6461 (×0.01)
°F
dTd
0011…0012
0262 (×0.01)
°C
6411…0012
6462 (×0.01)
°F
Tw
0013…0014
0263 (×0.01)
°C
6413…0014
6463 (×0.01)
°F
a
0015…0016
0264 (×0.01)
g/m3
6415…0016
6464 (×0.01)
gr/ft3
x
0017…0018
0265 (×0.01)
g/kg
6417…0018
6465 (×0.01)
gr/lb
h
0019…0020
0266 (×0.01)
kJ/kg
6419…0020
6466 (×0.01)
btu/lb
Available measurements depend on the instrument configuration. Values
may be unavailable also in case of device failure. Read status registers or
exception status outputs to check for failures.

Table of Contents

Related product manuals