Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 53
Set Output Format
Use the serial line command FORM to change the measurement message
sent by the transmitter on the service port. You can freely define the
output message to include the desired parameters, formatting options,
text strings, and additional fields.
FORM [modifier string]<cr>
where
modifier string
=
String of parameters and modifiers that defines the
output format, length 1 ... 15
0 characters.
Maximum length may be shorter when text strings
are used. See
Table 9 and Table 10 on page 54,
and examples below.
Command to set default format:
>form /
Example of default output:
RH = 5.17 %RH T = 24.33 'C
RH = 5.17 %RH T = 24.33 'C
RH = 5.18 %RH T = 24.33 'C
RH = 5.18 %RH T = 24.33 'C
...
Command to set output format as Tdf and T with Modulus-256
checksum:
>form "Tdf =" U3 4.2 tdf " T =" U3 3.2 t CS2 \r \n
Output example:
Tdf = -15.72 'C T = 24.38 'C C9
Tdf = -15.71 'C T = 24.38 'C C8
Tdf = -15.71 'C T = 24.38 'C C8
Tdf = -15.69 'C T = 24.38 'C CF
...
Command to set output format as Tdf and T, with start of text (ASCII
character 002) and end of text (003) ASCII codes, and without line feed
and carriage return at the end:
>form #002 "Tdf =" U3 3.2 tdf " T =" U3 3.2 t #003
Output example (ASCII codes not visible here):
Tdf =-15.14 'C T = 24.40 'CTdf =-15.14 'C T = 24.40 'CTdf
=-15.14 'C T = 24.40 'CTdf =-15.14 'C T = 24.40 'CTdf =-
15.13 'C T = 24.40 'CTdf =-15.13 'C T = 24.40 'C ...