Wiring
Read the product documentation carefully before installing or
operating the product. If you encounter the following marking during
installation or operation, consult product documentation to find out the nature
of the potential hazards and any actions which have to be taken to avoid them:
WARNING!
Make sure that you prepare and connect only de-energized wires.WARNING!
• Allen key (4 mm) for opening the transmitter cover
• Open-ended wrenches of sizes 17 mm, 19 mm, 22 mm and 24 mm
• Flat head screwdriver
• Cable glands as required by your application (available from Vaisala)
• Cables as required by your application (probe connection cables and power
cables are available from Vaisala)
Power supply terminals and lead-through − PELV
option
Attach only cables with temperature rating of minimum +80 °C
(+176 °F) to the PELV power supply terminal.
WARNING!
ANALOG OUTPUTS
RELAY 1RELAY 2
+
+
+
+
+
+
+
+
+
CH1
CH2
CH3
CH4
RS-485
PROBE
24 VOUT
ANALOG
INPUT
NO COM NC NO COM NC
1
2
M20×1.5
VS
+
Power supply input wiring is required only
for transmitters powered with protective
extra-low voltage (PELV) or AC (mains)
power.
For the M20×1.5 cable gland ordered
together with the transmitter from Vaisala,
the cable diameter is 5.0 … 8.0 mm
(0.20 … 0.31 in). Tightening torque for the
cable gland is 8 Nm.