User’s Guide ______________________________________________________________________
106 _________________________________________________________________ M211840EN-C
Command format: aCx!
The response:
atttnn<cr><lf>
where
a = Device address
C = Start concurrent measurement command
x = The desired measurement
1 = Wind
2 = Temperature, humidity, and pressure
3 = Precipitation
5 = Supervisor
If x is left out, the query refers to combined data
message in which the user can request data from
several sensors with just one command. See the
examples below.
! = Command terminator
where
a = Device address
ttt = The measurement completing time in seconds
nn = The number of the measured parameters available
(maximum number is 20)
<cr><lf> = Response terminator
NOTE
For information on changing the message parameters, units and other
settings, see to Chapter 8 Sensor and Data Message Settings, on page
127.