EasyManua.ls Logo

Valmet TS - Temperature Compensation Correction

Valmet TS
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Temperature compensation correction
The Valmet TS compensates for the effect of temperat-
ure by using a temperature measurement. If temperat-
ure sensitivity is still observed in the measurement
result, it can be eliminated by temperature compensa-
tion correction. First determine the actual temperature
dependence by means of laboratory monitoring. To
ensure reliable results, both the sampling and laborat-
ory analyses must be accurate and reliable, and also
the temperature range must be wide enough.
The results of laboratory monitoring are shown in a
TS-Lab vs. Temp graph. If the TS-Lab points are on a
horizontal graph, the Valmet TS measurement results
show no dependence on temperature. If the graph is
horizontal but are not at zero level, perform a level
correction to the calibration as instructed in section 7.3.
If the graph is not horizontal, the Valmet TS meas-
urement results show some dependence on temperat-
ure (Fig. 1). To eliminate this effect, enter a correction
curve for the temperature compensation. Select "Spe-
cial functions" -› "Temperature comp", and the following
display will appear:
T e m p T S - L a b
1 . 0 . 0 0 0 . 0 0
2 . 0 . 0 0 0 . 0 0
3 . 0 . 0 0 0 . 0 0
4 . 0 . 0 0 0 . 0 0
5 . 0 . 0 0 0 . 0 0
6 . 0 . 0 0 0 . 0 0
Press "Edit/Save". Then enter the correction curve
as point pairs, Temp/TS-Lab. The Valmet TS will use
this data to correct the compensation. Up to six point
pairs can be entered. In the case shown in Fig. 1 two
point pairs are enough, and based on the results these
can be selected for example as follows:
T e m p T S - L a b
1 . 4 2 . 0 0 . 0 0
2 . 4 8 . 1 0 . 1 0
3 . 0 0 . 0 0 . 0 0
4 . 0 0 . 0 0 . 0 0
5 . 0 0 . 0 0 . 0 0
6 . 0 0 . 0 0 . 0 0
Compensation is also successful outside the defined
interval such that the compensation curve is continued
beyond the outermost points using the same slope.
TS-Lab vs Temp
-0,2
-0,15
-0,1
-0,05
0
0,05
0,1
0,15
0,2
41 42 43 44 45 46 47 48 49 50
Temp (°C)
TS-Lab(%)
Fig. 1. Determining temperature dependence.
Installation & Owner's manual, OUL00489 V2.3 EN
28
Valmet Total Solids Transmitter

Table of Contents