47
6 Set up with the display and adjustment module
VEGAPULS 64 • Two-wire 4 … 20 mA/HART
51141-EN-210219
In the menu item " Current output Min./Max.", you determine the reac-
tion of the current output during operation.
The default setting is min. current 3.8 mA and max. current 20.5 mA.
In the menu item " Lock/unlock adjustment" you safeguard the sensor
parametersagainstunauthorizedorunintentionalmodications.
With active PIN, only the following adjustment functions are possible
without entering a PIN:
•
Select menu items and show data
•
Read data from the sensor into the display and adjustment module
Releasing the sensor adjustment is also possible in any menu item by
entering the PIN.
Caution:
With active PIN, adjustment via PACTware/DTM and other systems is
also blocked.
This menu item enables the setting of the requested national lan-
guage.
The following languages are available:
•
German
•
English
•
French
•
Spanish
•
Russian
•
Italian
•
Dutch
•
Portuguese
•
Japanese
•
Chinese
•
Polish
•
Czech
•
Turkish
In the delivery status, the VEGAPULS 64 is set to the ordered national
language.
Setup - Current output
Min./Max.
Lock/unlock setup - Ad-
justment
Display - Menu language