106
14 Diagnosis, asset management and service
VEGAPULS 6X • Two-wire 4 … 20 mA/HART
66190-EN-221107
Code
Text message
Cause Rectication DevSpec
State in CMD 48
F036
No operable soft-
ware
Failed or interrupted software up-
date
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Byte 5, Bit 3 of
Byte 0 … 5
F040
Error in the elec-
tronics
Hardware defect Exchanging the electronics
Send instrument for repair
Byte 5, Bit 4 of
Byte 0 … 5
F080
General software
error
General software error Disconnect operating voltage
briey
Byte 5, Bit 5 of
Byte 0 … 5
F105
Determine meas-
ured value
The instrument is still in the
switch-on phase, the measured
value could not yet be determined
Wait for the end of the switch-on
phase
Duration up to approx. 3 minutes
depending on the version and pa-
rameter settings
Byte 5, Bit 6 of
Byte 0 … 5
F113
Communication
error
EMC interference RemoveEMCinuences Byte 4, Bit 4 of
Byte 0 … 5
F125
Impermissible elec-
tronics temperature
Temperature of the electronics in
thenon-speciedrange
Check ambient temperature
Insulate electronics
Use instrument with higher tem-
perature range
Byte 5, Bit 7 of
Byte 0 … 5
F260
Error in the cali-
bration
Error in the calibration carried out
in the factory
Error in the EEPROM
Exchanging the electronics
Send instrument for repair
Byte 4, Bit 0 of
Byte 0 … 5
F261
Error in the instru-
ment settings
Error during setup
False signal suppression faulty
Error when carrying out a reset
Repeat setup
Carry out a reset
Byte 4, Bit 1 of
Byte 0 … 5
F264
Installation/Setup
error
Adjustment not within the vessel
height/measuring range
Max. measuring range of the in-
strumentnotsucient
Check or correct installation and/
or parameter settings
Use an instrument with bigger
measuring range
Byte 4, Bit 2 of
Byte 0 … 5
F265
Measurement func-
tion disturbed
Sensor no longer carries out a
measurement
Operating voltage too low
Check operating voltage
Carry out a reset
Disconnect operating voltage
briey
Byte 4, Bit 3 of
Byte 0 … 5
F267
No executable sen-
sor software
Sensor cannot start Exchanging the electronics
Send instrument for repair
-