EasyManua.ls Logo

Vega VEGAPULS 6X - Exchanging the electronics module

Vega VEGAPULS 6X
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
118
14 Diagnosis, asset management and service
VEGAPULS 6X • Two-wire 4 … 20 mA/HART
66190-EN-221107
Fault description Cause Rectication
Measured value
uctuatesaround
10 … 20 % (only bulk
solids)
Level
Various echoes from an uneven medium
surface, e.g. an extraction funnel
Check parameter "Type of medium" and
adapt, if necessary.
Optimize installation position and sensor
orientation.
Reectionsfromthemediumsurfaceviathe
vesselwall(deection)
Depending on the reason for the fault and the measures taken, the
steps described in chapter " Setup" must be carried out again or must
be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside normal working hours, seven
days a week around the clock.
Sinceweoerthisserviceworldwide,thesupportisprovidedin
English. The service itself is free of charge, the only costs involved are
the normal call charges.
14.6 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user.
In Ex applications, only instruments and electronics modules with ap-
propriate Ex approval may be used.
If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron-
icsmodulesareadaptedtotherespectivesensoranddierinsignal
output or voltage supply.
The new electronics module must be loaded with the default settings
of the sensor. These are the options:
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial
numbers are stated on the type label of the instrument, on the inside
of the housing as well as on the delivery note.
Whenloadingonsite,theorderdatamustrstbedownloadedfrom
the Internet (see operating instructions " Electronics module").
Information:
Allapplication-specicsettingsmustbeenteredagain.That'swhyyou
have to carry out a fresh setup after exchanging the electronics.
Ifyousavedtheparametersettingsduringtherstsetupofthesen-
sor, you can transfer them to the replacement electronics module. A
fresh setup is then not necessary.
Reaction after fault recti-
cation
24 hour service hotline

Table of Contents

Other manuals for Vega VEGAPULS 6X

Related product manuals