PG PLUS RACK + ZERO AIR
& GC PLUS RACK USER MANUAL
www.vicidbs.com
Version 2.00.0000
The generator will rst build the internal pressure. Wait until it
reaches 100% before moving on to the next step.
At this point, make sure that you have connected the outlet
Hydrogen to your GC or application. Once done, press Open.
The hydrogen will go through the outlet and the pressure will rise
until it reaches the selected pressure.
Once reached, the generator will show “Normal Flow”
7.7. STARTING THE ZERO AIR GENERATOR
The zero air generator needs to be connected to an external compressed air source.
In order to protect the generator, it is necessary to place a specic dust lter at the air inlet of the generator
(delivered with the generator).