8.5. SPECIAL FUNCTIONS
SELECTION DESCRIPTION
System test Select a test function below
Internal leak test The outlet valve is closed, the pressure is set to the max., when the pressure
reaches the max., production is stopped and the pressure drop is measured
over 2 minutes. The pressure drop should below 0.7 bars or 10 PSI
Important: The leak test will only work if the generator is in standby mode.
External leak test Works similar to the leak internal except the outlet valve is open and the test
pressure is equal to the set pressure. This function can be used to test the
external gas lines.
Important: The leak external will only work if the generator is in standby mode.
Flow test This functions sets the outlet valve to provide a certain ow adjustable from
0 to 100% of maximum ow.
Important: This function will only work if the generator is in standby mode. It
will take approx. 2 minutes to get a stable ow.
Complete system test Combination of all tests above
Hardware options Not available for user (password protected)
9. OPERATION - ZERO AIR PART
9.1. CONNECT POWER AND START UNIT
To start the unit:
• Power up the unit using the power button “Air” of the unit.
• Verify that the unit is powered by checking the green LED “POWER” light.
9.2. WARMING UP
After powering the generator, please wait 45 minutes before providing compressed Air to it. The zero air
generator is designed to run continuously. Continuous operation preserves the performance of the catalyst.
Shutdown should only be performed for maintenance. After the warmup cycle is completed, verify that:
• “POWER” green LED is on
• “READY” LED is on