37
PG PLUS RACK + ZERO AIR
& GC PLUS RACK USER MANUAL
www.vicidbs.com
Version 2.00.0000
Low Ext. Pressure The external pressure cannot be hold Alarm Check for external leaks. Check
max. H
2
consumption.
Low Water Level There is not enough water in the tank Alarm Rell the tank.
Bad Water The conductivity of the water has ex-
ceeded 25µS/cm
Alarm Drain and rell the tank. Change
the deionizer bags.
High Cell Voltage High cell voltage Alarm Notify service agent, cell must be
replaced.
Water Flow Error Too less water ows through the cell and
ow meter
Alarm Check to make sure water is in the
tank. Call service to check if the cell
or owmeter is blocked.
PS Temp High Temperature on the main board reaches
70°C
Alarm Check to make sure all fans are
working and air stream is not
blocked.
Dryer Failure Communication failure between
dryer module and main board
Alarm Check cable.
Shock/Fall The G-force sensor is activated due too
high force example earthquake
Alarm Check connections, tubing, etc.
before restarting the generator.
10.2. ZERO AIR PART
10.2.1. Front Panel LED
1. Green LED “Power” off
• Verify power cord connection.
• Verify that you are using the proper fuses and power supply voltage settings.
2. Red LED “Check System” on
Red LED is on if the internal heater has failed and the thermocouple reads a value higher than the allowed
maximum of 704°C (1300 °F). In this case, the power supply to the heater is stopped.
• Turn the unit off and on again to reset to initial status.
• If the problem cannot be solved with these steps, contact your local distributor.