56 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
Sterilization cycles
1 On the homepage, tap the cycle and .
2 Tap the leaf icon to set the ECO DRY mode.
3 Tap the ECO DRY mode you desire.
4 Tap to confirm and go back to the previous page.
5
To change the start time, tap Start cycle at.
6 Tap the time or the date: a settings page opens.
7 Tap the number you want to change and tap or to
increase it or decrease it.
8 Tap to confirm and go back to the previous page. This date
and time become the scheduled default start time for all next
sterilization cycles.
9 Tap to lock the door and start the countdown for the
current cycle.
Note: You can decide to start the cycle immediately or stop the
countdown at any time.
PROGRAM A SEQUENCE OF CYCLES TO RUN AUTOMATICALLY
It is possible to program the sterilizer in such a way that a sequence
of cycles runs automatically at a preset date and time (e.g. if you
want to run a vacuum test in the early morning before office hours).
Maximum two cycles can be programmed.
To program the sterilizer proceed as follows:
1 Load the chamber and close the chamber door.
2 On the homepage, tap > > >
3 Tap to add a cycle to the list above.
4 Tap and then the cycle you want to add.
5 Tap : the cycle is automatically listed in the previous page.
6 Tap the date/time to set the start of the cycle.
7 Tap the number you want to change and tap or to
increase it or decrease it.
8 Tap to confirm and go back to the previous page.
9 Repeat steps from 3 to 7 to add another cycle to the
sequence.
10 Tap to launch the programmed sequence.
DELETE A CYCLE FROM A PROGRAMMED SEQUENCE
1 On the homepage, tap > > >
2 Tap the cycle you want to remove from the sequence.
3 Tap to remove the selected cycle.
4 Tap to confirm and go back to the previous page.