Environmental specifications B-3
Environmental specifications
Environmental specifications
Item Specification
Operating temperature 4 to 40 °C
Humidity 20 to 80%, noncondensing
Acoustic noise
(instrument-generated)
<60 dBA at idle
Solvent compatibility Solvents consistent with materials of
construction. Salts and buffers can reduce seal
life, especially at pressures in excess of
20,700 kPa (207 bar, 3000 psi).
Internal cooling DC-powered fans provide internal cooling by
pulling air through the binary gradient module
and exhausting it out the rear of the unit.
Airflow requirements Air enters the front of the binary gradient
module so pumps can be placed side by side.
Clearance Allow at least 6 inches (15 cm) between the rear
of the binary gradient module and the wall for
power connections and the ventilation fans’
exhaust.