B-8 Specifications
Composition accuracy ±1.0% absolute (full scale) from 5 to 95% from
1.0 to 130 mL/min using water/water or
MeOH/MeOH spiked with dye or marker, back
pressure approximately 4140 kPa (41 bar,
600 psi).
Composition precision Retention time variation less than 0.3% RSD or
0.1 min SD, whichever is greater. Standard
tests at 1.00, 20.00, 50.00 mL/min, test mix
Uracil/Caffeine, Xterra columns 4.6 × 50, 19 ×
50, 30 × 50, injection volumes 6, 100, 500 µL.
Ambient temperature fluctuations may affect
results.
Gradient composition
curves
Eleven gradient curves (including linear, step
[2], concave [4], and convex [4]).
Solvent management system specifications (Continued)
Item Specification