Creating the inlet method 2-9
3. In the Solvent Names area, enter names for Solvent A and Solvent B. To
change solvent names, enter a name for solvent A in the label A box and
enter a name for Solvent B in the label B box.
Examples: acetonitrile, methanol, water.
4. In the Pressure Limits area, enter the low-pressure limit for the system
in the Low field. If the system pressure falls below this limit, the flow
stops and the LC Status error light turns red.
Range: 0 to 41,400 kPa (414 bar, 6000 psi). The default is 0 psi.
5. Enter the high-pressure limit for the system in the High field. If the
system exceeds this limit, the flow stops and the LC Status error light
turns red.
Range: 0 to 41,400 kPa (414 bar, 6000 psi). The default is 6000 psi.
6. In the Seal Wash Period field, enter the time interval between seal-wash
pump pulses. The seal wash lubricates the plunger and flushes away
solvent or precipitated salts forced past the plunger seal from the
high-pressure side of the solvent piston chamber.
Range: 0.5 to 9999 minutes. The default is 5 minutes.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters 2545 Binary Gradient Module and is the answer not in the manual?

Waters 2545 Binary Gradient Module Specifications

General IconGeneral
BrandWaters
Model2545 Binary Gradient Module
CategoryControl Unit
LanguageEnglish

Related product manuals