EasyManua.ls Logo

Waters 2545 Binary Gradient Module - Page 9

Waters 2545 Binary Gradient Module
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Features 1-3
The two independent pump systems combine their two solvents at a filter/tee,
where the solvent mix is plumbed to the sample manager or another
component of the AutoPurification System.
MassLynx software controls the ratio of the two solvents by varying the
relative flows of pump A and pump B. The binary gradient module receives
pressure data from a pressure transducer associated with each pump head.
The firmware measures the individual pump head pressure during the
pumping cycle and adjusts the precompression to ensure consistent solvent
delivery and minimize pump-induced detector baseline disturbances.
The MassLynx interface with the binary gradient module also provides useful
diagnostics and troubleshooting routines.
Leak management
All fluid-handling areas of the binary gradient module contain leak routing
capability. Spilled solvent is routed to the drain tube below the front panel.
The optional leak sensor should be installed on the leak sensor bracket in the
drip tray.
See also: Waters AutoPurification System Guide and Waters System Fluidics
Organizer Operator’s Guide.

Related product manuals