Replacing the IonSABRE APCI probe sample capillary 5-65
Installing the new capillary
Required materials:
• Chemical-resistant, powder-free gloves
• Needle-nose pliers
•7-mm wrench
•8-mm wrench
•2.5-mm Allen wrench
• Length of red PEEK tubing
•LC pump
• HPLC-grade, or better, 1:1 acetonitrile/water
• Replacement capillary
• Replacement ferrule