2-4 Preparing for Operation
Configuring IntelliStart
To configure IntelliStart
1. In the console system tree, expand 3100 Detector.
2. Click IntelliStart.
3. Click Configure > IntelliStart Configuration.
4. In the IntelliStart Configuration dialog box, in the Checks list, select the
check boxes for the items you want checked during 3100 Detector
startup. Clear the check boxes of items you do not want checked.
Tip: To display detailed information for an item, highlight it and then
click Properties.
5. Click OK.
Verifying the instruments state of readiness
When the instrument is in good operating condition, the power and Operate
LEDs show constant green. You can view any error messages in IntelliStart.
To access IntelliStart
1. In the console system tree, expand 3100 Detector.
2. Click IntelliStart.
Tuning and calibration information
The instrument must be tuned and calibrated prior to use, tasks normally
performed from IntelliStart.
For further instruction, see the instrument’s online Help topic “Instrument
Setup”.