Cleaning the sample cone and gas cone 5-27
Fitting the cone gas assembly to the source
Required material: Chemical-resistant, powder-free gloves
To fit the cone gas assembly to the source
Warning: The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning: To avoid puncture wounds, take great care while working
with the source enclosure door open if one or both of these conditions
apply:
An ESI probe is fitted (the probe tip is sharp).
A corona pin is fitted (the pin tip is sharp).
Caution: To avoid damage,
do not apply any downward force to the source enclosure door while
the door is open.
do not open the source isolation valve before fitting the cone gas
assembly to the ion block assembly.