Replacing the source assembly seals 5-79
Disassembling the source enclosure and probe adjuster
assembly
Required materials:
Chemical-resistant, powder-free gloves
4-mm Allen wrench
To dismantle the source enclosure and probe adjuster assembly
1. Use the 4-mm Allen wrench to remove the 4 screws securing the probe
adjuster assembly to the source enclosure.
2. Remove the probe adjuster assembly from the source enclosure.
3. Unscrew and remove the 3 thumbscrews and door spring-clip that
secure the source enclosure side flange to the source enclosure.
4. Remove the source enclosure side flange from the source enclosure.
Warning: The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
RP00024
Source enclosure
side flange
Thumbscrew
Probe adjuster
assembly
Probe adjuster assembly
securing screw