Emptying the exhaust trap bottle 5-11
To empty the exhaust trap bottle
1. In the console, click Stop Flow to stop the LC flow.
2. In the console, click API to stop the desolvation gas flow.
3. Unscrew and remove the exhaust trap bottle cap.
4. Dispose of the waste liquid in accordance with local environmental
regulations.
5. Fit and tighten the exhaust trap bottle cap.
6. Ensure that the exhaust trap bottle is secured in the upright position.
7. In the console, click API to start the desolvation gas flow.
8. In the console, click Start Flow to start the LC flow.
9. Use snoop (or equivalent) leak detector liquid to ensure that there are no
leaks at the exhaust trap bottle cap.
Warning: The waste liquid in the exhaust trap bottle
comprises LC solvents and analytes. Always wear
chemical-resistant, powder-free gloves while handling the
exhaust trap bottle.
Warning: The waste liquid can be contaminated with
biohazardous and/or toxic materials. Ensure that it is
correctly disposed of according to local environmental
regulations.
Warning: To confirm the integrity of the source exhaust
system, the following leak test must be performed.
Caution: To avoid damage to the instrument, snoop (or
equivalent) leak detector liquid must be used only for the
purpose described in the following step; it must not be used on
any other part of the instrument.