EasyManua.ls Logo

Waters ACQUITY QDa - Starting the Instrument

Waters ACQUITY QDa
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
June 28, 2016, 715003956 Rev. C
Page 23
Figure 2–1: Inserting the probe
3. Secure the probe tubing to the side of the source enclosure using the clip.
2.4 Starting the instrument
Starting the instrument entails powering-on the ACQUITY workstation, logging into the
workstation, powering-on the instrument and all other ACQUITY instruments, and starting the
software.
Requirement: You must power-on and log in to the ACQUITY UPLC workstation first to ensure
that it obtains the IP addresses of the system instruments.
Warning: To avoid electric shock and burn injuries, follow these guidelines:
Do not expose the data system or any ancillary equipment to dripping or splashing liquids.
Do not place objects filled with liquid, such as solvent bottles, on the data system or ancillary
equipment.
Notice: To avoid damage to the instrument caused by incompatible solvents, refer to
Appendix C, page 71.
Probe assembly
Probe fitting
PEEK fitting
Capillary

Table of Contents

Related product manuals