June 28, 2016, 715003956 Rev. C
Page 31
To remove the source components from the ion block:
1. Remove the source enclosure (see page 28).
2. To remove the gas cone and PEEK cone clamp, pull the cone clamp away from the
instrument.
Figure 3–2: Removing the gas cone and PEEK cone clamp
3. Remove the sample cone assembly from the right-hand side of the ion block, and
disassemble the parts by hand.
Notes:
Three parts compose the sample cone assembly: a metal sample cone; a disc aperture
assembly; a metal entrance-aperture carrier.
The disc aperture assembly contains a metal disc. If the disc becomes separated from
the rubber carrier, simply re-insert the disc into the carrier.
Requirement: The Standard and Performance QDa Detectors require different entrance
aperture discs. When ordering a replacement entrance-aperture disc and seal, ensure that
the entrance-aperture disc is the correct type.
Warning: To avoid burn injuries, allow the ion block to cool before performing this
procedure.
PEEK cone clamp
Ion block
Gas cone