Replacing the ESI probe tip and gasket 4-67
3. Remove the metal gasket from the probe tip.
4. Dispose of the metal gasket in accordance with local environmental
regulations.
5. If the probe tip is damaged, dispose of it in accordance with local
environmental regulations.
Fitting the ESI probe tip and gasket
Required materials
Chemical-resistant, powder-free gloves
•10-mm wrench
New metal gasket
Warning: The probe tip and metal gasket can be
contaminated with biohazardous and/or toxic materials.
Dispose of them according to local environmental
regulations.
Metal gasket