EasyManua.ls Logo

Waters Xevo TQD - Removing the ESI Probe

Waters Xevo TQD
250 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-4 Changing the Mode of Operation
6. Using PEEK tubing equal to 0.004-inch ID, connect port S of the selector
valve to the ESI probe.
Recommendation: To reduce peak broadening, use 0.004-inch ID tubing
for sample flow rates 1.2 mL/min; use 0.005-inch ID tubing for sample
flow rates 1.2 mL/min.
Requirements:
If you are replacing the tubing between the selector valve and the
probe, minimize the length to reduce peak broadening.
When cutting the tubing to length, cut it squarely (that is,
perpendicular to its horizontal axis).
Removing the ESI probe
Required material
Chemical-resistant, powder-free gloves
To remove the ESI probe:
1. Prepare the instrument for working on the source (see page 4-7).
2. Disconnect the tubing from the ESI probe.
3. Disconnect the ESI probe cable from the high-voltage connector.
4. Unscrew the probe locking ring.
Warning: To avoid electric shock, do not use stainless steel tubing
or stainless steel finger-tight screws to connect the selector valve
to the ESI probe; use the PEEK tubing and natural (beige) colored
PEEK finger-tight screws supplied with the instrument.
Warning: The LC system connections, ESI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning: To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.

Table of Contents

Other manuals for Waters Xevo TQD

Related product manuals