Combined APPI/APCI source 3-15
Removing the IonSABRE II probe and APPI/APCI source
enclosure
Required material
Chemical-resistant, powder-free gloves
To remove the combined APPI/APCI source
:
1. Prepare the instrument for working on the source (see page 4-7).
2. Remove the IonSABRE II probe (see page 3-8).
3. Disconnect the HT cable from the instrument’s front panel.
4. Disconnect the APPI drive cable from the instrument’s front panel.
5. Remove the source enclosure (see page 4-8).
6. Remove the corona pin (see page 4-15).
7. Fit the blanking plug to the pin’s mounting contact.
Warning: The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning: To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
Warning: The probe and source can be hot. To avoid burn injuries,
take great care while working with these components.