NanoFlow source 3-19
9. Connect the high-voltage cable to the instrument’s HV connection.
10. Slide closed the instrument’s source interface door.
Fitting a borosilicate glass capillary (nanovial)
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
Borosilicate glass capillary
Fused silica syringe needle or GELoader
®
tip
Fused silica cutter
Caution: The NanoFlow stage contains a high-voltage interlock, so
the capillary voltage (the voltage applied to the sprayer assembly)
and the sampling cone voltage remain disabled until the sprayer is
pushed fully forward in the source.
High-voltage cable