4-114 Maintenance Procedures
19. Use the O-ring removal kit to carefully remove the shaft mounting
flange O-ring from the APPI source enclosure side.
20. Dispose of the O-rings in accordance with local environmental
regulations.
Warning: The O-rings can be contaminated with biohazard-
ous and/or toxic materials. Ensure that they are correctly
disposed of according to local environmental regulations.
Lamp mounting flange O-ring