1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter 6345 as the Advanced settings code.
d. Touch OK.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
2. Touch the Network tab.
The Status tab appears at the top of the screen.
3. Touch the vertical Radio or Ethernet tab to view the wireless Radio or Ethernet IP,
MAC address, and Status information.
4. Touch (Next) to view more Ethernet or Radio information settings.
Information in the Status tab updates only when the device is connected to a wired
or wireless network.
5. Do one of the following:
• To continue in the Advanced Settings, touch another tab.
• To exit the Advanced Settings and return to the Home screen, touch Exit.
Specify Ethernet settings
You can connect a CP150 electrocardiograph to a Welch Allyn CardioPerfect workstation
or a network server via an Ethernet cable. Software provided in the CP 50/150
connectivity kit is required to communicate with the workstation. Cables longer than 3
meters have not been verified for use with the electrocardiograph. Do not use cables
longer than 3 meters.
1.
Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter 6345 as the Advanced settings code.
d. Touch OK.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
If connecting the CP150 electrocardiograph to a Welch Allyn CardioPerfect
workstation, connect the electrocardiograph to the same network as the
workstation. If you need help, consult your network administrator.
2. Touch the Network tab.
3. Touch the Ethernet tab.
Service manual Settings 31