EasyManua.ls Logo

wtw Multi 9420 - Measuring the Temperature; FDO Check Procedure; Why Should You Check the Sensor; When Should You Check the Sensor

wtw Multi 9420
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi 9420 Dissolved oxygen (D.O.)
ba75964e04 04/2015 71
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
7.1.2 Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
7.2 FDO
®
Check procedure
7.2.1 Why should you check the sensor?
With the FDO
®
Check procedure, you can find out in a simple manner
whether the FDO
®
925 D.O. sensor should be cleaned or calibrated.
7.2.2 When should you check the sensor?
Checking can be useful in the following cases:
When the check interval has expired
(the [check] status indicator is displayed).
If the measured values seem to be implausible
If you assume that the sensor cap is contaminated or at the end of
its lifetime
After the sensor cap was exchanged
Routinely within the framework of the company quality assurance
7.2.3 Carrying out the FDO
®
Check procedure
FDO
®
Check
procedure
Check in water vapor-saturated air.
Use the check and storage beaker (FDO
®
Check) to carry out theFDO
®
Check
procedure.
Measured parameter Time interval Stability in the time
interval
D.O. concentration 20 seconds
: better than 0.03 mg/l
D.O. saturation 20 seconds
: better than 0.4 %
D.O. partial pressure 20 seconds
: Better than 0.8 mbar
Temperature 15 seconds
: Better than 0.5 °C

Table of Contents

Related product manuals