Multi 9420 Settings
ba75964e04 04/2015 95
9.6 Sensor-independent settings
9.6.1 System
To open the Storage & config menu, press the <ENTER_> key in the measured
value display. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.
Menu item Setting Explanation
System / General / Language Deutsch
English
(more)
Selects the menu language
System / General / Audio signal on
off
Switches on/off the beep on key-
stroke
System / General / Brightness 0 ... 10 ... 22 Changes the display brightness
System / General / Tempera-
ture unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahren-
heit.
All temperature values are displayed
with the selected unit.
System / General / Stability
control
on
off
Switches on or off the automatic sta-
bility control during measurement
(see section 9.6.3 A
UTOMATIC STABI-
LITY CONTROL, page 96 )
System / Interface / Baud rate 1200, 2400, 4800,
9600, 19200
Baud rate of the USB Device inter-
face
System / Interface / Output for-
mat
ASCII
CSV
Output format for data transmission
For details, see section 11 T
RANSMIT-
TING DATA, page 107
System / Interface / Decimal
separator
Dot (xx.x)
Comma (xx,x)
Decimal separator
System / Interface / Output
header
Output of a header for Output format:
CSV
System /Clock function Date format
Datum
Time
Settings of time and date.
For details, see section 3.8.5
E
XAMPLE 2 ON NAVIGATION: SETTING
THE DATE AND TIME, page 25
System /Service information Hardware version and software ver-
sion of the meter are displayed.
System /Reset - Resets the system settings to the
default values.
For details, see section 9.7.2 R
ESET-
TING THE SYSTEM SETTINGS, page 100

Table of Contents