Settings Multi 9420
94 ba75964e04 04/2015
Setting menu of
LR 925/01
Menu item Possible
setting
Explanation
Cell constant 0.090 ... ...
0.100 ...
0.110 cm
-1
Display and setting options for the
cell constant
Temp. comp. (TC) /
Method
nLF
lin
off
Procedure for temperature compen-
sation (see section 8.2 T
EMPERA-
TURE COMPENSATION, page 80).
This setting is only available for the
measured parameters, ϰ and ρ.
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available
when the linear temperature com-
pensation is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters, ϰ and ρ.
Multiplier for TDS 0.40 ... 1.00 Factor for TDS value
Reset - Resets all sensor settings to the
delivery condition (see section 9.7.1
R
ESETTING THE MEASUREMENT SET-
TINGS, page 97).