Ion concentration Multi 9420
66 ba75964e04 04/2015
6.3.5 Standard addition with blank value correction (Blank value
addition)
In the "Standard addition with blank value correction" procedure, a known
amount of standard solution is added to the sample in two steps.
With the first addition, the ion concentration is raised to the linear range of the
electrode characteristic curve.
The second addition is equivalent to the standard addition. The ion concentra-
tion in the sample is calculated from the change in potential.
Calibration
17. If necessary, start measuring further samples with <ENTER>.
Repeat steps 8 - 16 for all samples.
18. Terminate the measuring method with <M>.
A safety query appears.
19. Select yes with <><>.
20. Confirm yes with <ENTER>.
Measurement with the selected method is completed.
1. Select the measuring method (see section 6.3 S
ELECTING THE MEASU-
RING METHOD, page 56).
01.08.2013 08:00
Sample subtraction
2. Prepare two standard calibration solutions.
3. Perform a two-point calibration according to user guidance.
4. The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.