EasyManua.ls Logo

Xylem Aquavar - Page 146

Xylem Aquavar
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Warning/Alarm Description Cause Remedy
48 — 1.8 V supply low The power supply is measured on
the control card.
The 1.8 V DC supply used on the
control card is outside of allowable
limits.
Check for a defective control card.
If an option card is present, check
for an overvoltage condition.
49 — Speed limit When the speed is not within the
specified range in [4–11] Motor
Speed Low Limit [RPM] and [4–13]
Motor Speed High Limit [RPM],
the frequency converter will show
a warning.
When the speed is below the
specified limit in [1–86] Trip
Speed Low [RPM] (except when
starting or stopping) the frequency
converter will trip.
50 — AMA calibration failed Contact your Xylem supplier or
Xylem Service Department.
51 — AMA check Unom and Inom The settings for motor voltage,
motor current, and motor power
are wrong.
Check the settings in [1–20] to [1–
25].
52 — AMA low Inom The motor current is too low. Check the setting in [4–18]
Current Limit.
53 — AMA motor too big The motor is too big for the AMA to
operate.
54 — AMA motor too small The motor is too small for the AMA
to operate.
55 — AMA Parameter out of range The parameter values of the motor
are outside of the acceptable
range. AMA will not run.
56 — AMA interrupted by the user The AMA has been interrupted by
the user.
57 — AMA timeout Try to restart AMA again. Repeated
restarts may overheat the motor.
58 — AMA internal fault Contact your Xylem supplier.
59 — Current limit The current is higher than the
value in [4–18] Current Limit.
Ensure that motor data in [1–20]
through [1–25] are set correctly.
Possibly increase the current limit.
Be sure that the system can
operate safely at a higher limit.
60 — Pump Protect (North
America)/External Interlock
(International)
A digital input signal is indicating
a pump protection external to the
controller is active.
62 — Output frequency at
maximum limit
The output frequency has reached
the value set in [4–19] Max Output
Frequency.
Check the application to
determine the cause. Possibly
increase the output frequency. Be
sure the system can operate safely
at a higher output frequency. The
warning will clear when the output
drops below the maximum limit.
65 — Control card over
temperature
The cut-out temperature of the
control card is 80°C.
Check that the ambient operating
temperature is within limits.
Check for clogged filters.
Check fan operation.
Check the control card.
7 Warnings and alarms
144 Aquavar
®
Intelligent Pump Controller - 150 HP to 600 HP INSTRUCTION MANUAL

Table of Contents