EasyManua.ls Logo

Xylem wtw Multi 3410 - Page 51

Xylem wtw Multi 3410
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Multi 3410
51
ba75863y18 01/2015
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
7 Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The pH value of the buffer solution is displayed.
8 Set the nominal buffer value for the measured temperature with
<><>.
9 Accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
10 If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.

Table of Contents

Related product manuals