4-5
5. If the sipper is not centered, touch [Position] and use the arrow buttons to center the sipper.
6. Make certain the Sipper is centered, then touch [Save] at the bottom of the adjustment window.
Sipper Adjustment Position
Figure 4-2
7. To test the alignment of the sipper, Touch [Inject] to lower the sipper, then touch [Retract] to raise the sipper back
up. If necessary, touch [Position] and repeat the adjustment.