EasyManua.ls Logo

Xylem YSI 2500 - Page 79

Xylem YSI 2500
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10-5
POSSIBLE CAUSE: Newly installed enzyme Membrane.
ACTION: Enter probe service and check probe currents.
SECTION: 6.2.2 Module
POSSIBLE CAUSE: Newly installed probe.
ACTION: Enter probe service and check probe currents.
SECTION: 6.2.2 Module
POSSIBLE CAUSE: Calibrator solution out of spec: contaminated or installed
for more than 30 days.
ACTION: Install new calibrator.
SECTION: 4.6 Install Calibrator Solution
POSSIBLE CAUSE: Failing enzyme Membrane.
ACTION: Enter probe service and check probe currents. Replace Membrane(s) if
necessary.
SECTION: 5.1 Perform Daily Operational Checks, 4.7 Install Enzyme Membranes
POSSIBLE CAUSE: Probe surface fouled.
ACTION: Clean probe surface.
SECTION: 8.4.3 Enzyme Probe Cleaning
POSSIBLE CAUSE: Auxiliary electrode fouled.
ACTION: Clean Auxiliary electrode surface (temperature probe) with isopropyl alcohol.
POSSIBLE CAUSE: Sample concentration too high, resulting in high probe
current (nA).
ACTION: Dilute sample or adjust sample size down and repeat.
SECTION: 7.1 Sample Volume
SYMPTOM: MEASUREMENT ERROR: PL Slope
POSSIBLE CAUSE: Stir speed too fast or too slow.
ACTION: Adjust stir speed.
SECTION: 6.2.2.4 Controls, Stirbar
POSSIBLE CAUSE: Newly installed enzyme Membrane.
ACTION: Enter probe diagnostics and check probe currents.
SECTION: 6.2.2 Module
POSSIBLE CAUSE: Newly installed probe.
ACTION: Enter probe diagnostics and check probe currents.
SECTION: 6.2.2 Module
POSSIBLE CAUSE: Failing enzyme membrane.
ACTION: Replace membrane.
SECTION: 5.1 Perform Daily Operational Checks, 4.7 Install Enzyme Membranes
POSSIBLE CAUSE: Enzyme Probe surface fouled.
ACTION: Clean probe surface.
SECTION: 8.4.3 Enzyme Probe Cleaning
POSSIBLE CAUSE: Auxiliary electrode fouled.
ACTION: Clean Auxiliary electrode surface (temperature probe) with isopropyl alcohol.
SYMPTOM: FLUID ERROR: Check Bottles
POSSIBLE CAUSE: Low buffer or calibrator level.
ACTION: Check bottle status at bottom of Sample screen. Refill the appropriate bottle.
SECTION: 4.5 Prepare and Install Buffer Solution, 4.6 Install Calibrator Solution

Table of Contents

Related product manuals