EasyManuals Logo

YOKOGAWA GX10 User Manual

YOKOGAWA GX10
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
2-140
IM 04L51B01-17EN
Detail
*
Meaning
BAD_ADDR
The address of the response message is invalid (does not match
the command).
BAD_NUM
The register of the response message is invalid (does not match
the command).
BAD_CNT
The number of registers in the response message is invalid (does
not match the command).
NO_DATA
Data has not yet been received once.
BAD_DATA
Data conversion of the response message failed.
VALID
Data is being acquired normally.
DROP_OUT
Communication dropout occurred due to the inability to keep up.
STALE
The response from the connected device is slow relative to the read
cycle.
START
Modbus or communication settings were changed.
STOP
Modbus or communication settings were changed.
* “_” expresses an underscore.
2.10.16 FTP Client Log (FLog)
The output in response to the command “FLog,FTP” is shown below.
Syntax
EA<crlf>
yyyy/mo/dd_hh:mm:ss_xxxxxxxxx_k_fff…f<crlf>
EN<crlf>
yyyy/mo/dd_hh:mm:ss
Time of error occurrence
yyyy
Year (1900 to 2099)
mo
Month (01 to 12)
dd
Day (01 to 31)
hh
Hour (00 to 23)
mm
Minute (00 to 59)
ss
Second (00 to 59)
xxxxxxxxx
Detailed code (fixed to 9 characters)
TCPIP_¯_¯
Internal processing error
HOSTADDR_
IP address not set
HOSTNAME_
Unable to resolve server host name
UNREACH_¯
Unable to connect to server
CONNECT_¯
Unable to connect to data port
SEND_¯_¯_
Transmission to data port failed
RECV_¯_¯_
Reception from data port failed
REPLY_¯_¯
Received reject response from server
SERVER_¯_
Invalid server response
CMDSEND_¯
Error in sending command to control port
CMDRECV_¯
Error in receiving command from control port
USER_¯_¯_
Invalid user name
PASS_¯_¯_
Invalid password
ACCT_¯_¯_
Internal processing error
TIMEOUT_¯
Response timeout
LINK_¯_¯_
Ethernet cable not connected
FILE_¯_¯_
File access failed
NOFD_¯_¯_
Internal processing error
NOID_¯_¯_
Internal processing error
PARAM_¯_¯
Internal processing error
CERT_¯_¯_
Certificate verification error
SSL_¯_¯_¯
Encryption communication error
k
Server type (P, S)
fff…f
File name (fixed to 51 characters including extension. Unused
character positions are filled with spaces.)
2.10 ASCII Output Format

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA GX10 and is the answer not in the manual?

YOKOGAWA GX10 Specifications

General IconGeneral
BrandYOKOGAWA
ModelGX10
CategoryVoice Recorder
LanguageEnglish

Related product manuals