2-151
IM 04L51B01-17EN
Commands and Responses
2
Operation Event string ### information and detailed information
Timer reset TimerRST Timer:ttt,ttt,ttt•••
ttt List of timer numbers that
were reset (ALL for all timers)
Match time timer
reset
MTimerRST Timer:ttt,ttt,ttt•••
ttt List of timer numbers that
were reset (ALL for all timers)
Communication
channel write
(screen operation
only)
WriteComm kk•••,CCCC=dd•••
kk••• Write type [Internal,External]
C Communication channel
dd••• Value
Example:
Internal,C001=1.234
DO channel write (for
manual operation)
(screen operation/
general
communication
command only)
WriteDO CCCC=dd•••
C DO channel
dd••• Value [ON, OFF]
Example:
0901=OFF
SW channel write
(for manual
operation) (screen
operation/general
communication
command only)
WriteSW CCCC=dd•••
C Internal switch
dd••• Value [ON, OFF]
Example:
S001=ON
Settings saved Save######### ##•# Save type (output in the event string)
Report Report
Scale Scale image
When multi batch is enabled,
a hyphen followed by the
specified batch group number
is added.
Example:
SaveScale-02
Custom Custom display
Parameter Setting parameter
Cert Certificate
All All settings
<Detailed information>
---- When ### = Report ----
cc•••,rr•••
cc••• Report format [EXCEL, PDF]
rr••• Report type [Hour, Day,
Week, Month, Hour+Day,
Day+Week, Day+Month,
Batch, Custom]
---- When ### = Scale ----
Group:gg
gg Group number
---- When ### = Custom ----
No:nn•••
nn••• Display number (ALL for all
custom display screen)
Settings loaded Load########## ##•# Load type (output in the event string)
Report Report
Scale Scale image
When multi batch is enabled,
a hyphen followed by the
specified batch group number
is added.
Example:
LoadScale-02
Custom Custom display
2.10 ASCII Output Format