2-160
IM 04L51B01-17EN
Output example when an expandable I/O or sub unit is connected
• The unit information area (e.g. U00) will contain the expansion module name. All seven
units are output regardless of whether expansion modules are available.
• The module information area (after Unit) will contain I/O module names. Only the units
that have modules installed will be output.
EA
U00*:GX20-1J GX20-1J ----------------
U01 :GX90EX-02-TP1 GX90EX-02-TP1 ----------------
U02 :---------------- ---------------- ----------------
U03 :---------------- ---------------- ----------------
U04 :---------------- ---------------- ----------------
U05 :---------------- ---------------- ----------------
U06 :---------------- ---------------- ----------------
Unit:00
00:GX90XA-10-U2 GX90XA-10-U2 ----------------
01:GX90XA-10-U2 GX90XA-10-U2 ----------------
02:---------------- ---------------- ----------------
03:---------------- ---------------- ----------------
04:---------------- ---------------- ----------------
05:---------------- ---------------- ----------------
06:---------------- ---------------- ----------------
07:---------------- ---------------- ----------------
08:---------------- ---------------- ----------------
09:---------------- ---------------- ----------------
Unit:01
00:GX90XA-10-U2 GX90XA-10-U2 ----------------
01:---------------- ---------------- ----------------
02:---------------- ---------------- ----------------
03:---------------- ---------------- ----------------
04:---------------- ---------------- ----------------
05:---------------- ---------------- ----------------
06:---------------- ---------------- ----------------
07:---------------- ---------------- ----------------
08:---------------- ---------------- ----------------
09:---------------- ---------------- ----------------
EN
cccccccccccccccc
Module models that are actually installed
-------------------- Module not installed (16 hyphens)
Displays the module model code.
uuuuuuuuuuuuuuuu
Module models recognized by the GX
-------------------- Module not installed (16 hyphens)
Displays the module model code.
defghijklmnopqrs
Module status
-
Normal
X
Error
d to s express the following items.
d
System data error
e
Calibration value error
f
Parameter error
g
Reserved (-)
h
FRAM error
i
Reserved (-)
j
Reserved (-)
k
Reserved (-)
l
A/D error
m
RJC error
n
Reserved (-)
o
Reserved (-)
p
Reserved (-)
q
Reserved (-)
r
Reserved (-)
s
Reserved (-)
nn
Unit number
f
* GX/GP or GM main unit
2.10 ASCII Output Format