13-11
IM 760401-01E
Communication Commands 1 (System of Commands before the IEEE 488.2-1987 Standard)
13
OY/OY? Sets the relay output items in case of
normal measurement/inquires about the
current setting. Up to four items can be
set.
Syntax OY m1,m2,m3,m4,m5 <terminator>
“m1” indicates the output relay
channel
1 m1 4
“m2” indicates the output item
number
m2= 0: no output
1: Voltage (V)
2: Current (A)
3: Power (W)
4: Reactive power (var)
5: Apparent power (VA)
6: Power factor (PF)
7: Input voltage frequency (V
Hz)
8: Input current frequency (A
Hz)
9: Watt-hour (Wh)
10: Ampere-hour (Ah)
11: Phase angle (deg)
12: Peak voltage value (Vpk)
13: Peak current value (Apk)
14: Computation result (MATH)
24: Positive watt-hour (Wh+)
25: Negative watt-hour (Wh–)
26: Positive ampere-hour (Ah+)
27: Negative ampere-hour (Ah–)
“m3” indicates element.
m= 1: Element 1
2: Element 2 (for WT230 model
760503 only)
3: Element 3 (for WT230 only)
4: (for WT230 only)
“m4” indicates setting value.
0.000 m4 ±9999
“m5” indicates prefix
m5= 0: m(E-3)
1: (E+0)
2: k(E+3)
3: M(E+6)
Query OY1? <terminator>
Example OY1,1,1,600.0,1
Description Parameter error 12 will occur if “m” is set to
an illegal value.
“No output” and “computation result” have no
relation to the element, but when using them
with the “OY” command, set m3=1.
OYH/OYH? Sets the relay output items in case of
harmonic measurement/inquires about
the current setting. Up to four items can
be set.
Syntax OYH m1,m2,m3,m4,m5,m6 <terminator>
“m1” indicates the output relay
channel
1 m1 4
“m2” indicates the output item
number
m2= 0: no output
1: Voltage (V)
2: Current (A)
3: Power (W)
6: Power factor (PF)
16: harmonic distortion factor
of voltage (V THD)
17: harmonic distortion factor
of current (A THD)
19: Relative harmonic content
of each voltage component
(V %)
20: Relative harmonic content
of each current component
(A %)
21: Relative harmonic content
of each active power
component (W %)
22: Phase angle between each
voltage of the 2nd to 50
(or 30) th order and the
fundamental (1st order)
voltage (V deg)
23: Phase angle between each
current of the 2nd to 50
(or 30) th order and the
fundamental (1st order)
current (A deg)
“m3” indicates element.
m= 1: Element 1
2: Element 2 (for WT230 model
760503 only)
3: Element 3 (for WT230 only)
“m4” indicates order of the
harmonic.
m4= Integer between 1 and 50 (or
30)
“m5” indicates setting value.
0.000 m5 ±9999
“m6” indicates prefix.
m6= 0: m(E-3)
1: (E+0)
2: k(E+3)
3: M(E+6)
Query OYH3? <terminator>
Example OYH3,3,1,1,1,200,2
Description Parameter error 12 will occur if “m” is set to
an illegal value.
“No output” is not related to any element,
order or setting value, so in case the OYH
command is set, set these all to 1 as a
dummy.
“PF”, “VTHD” and “ATHD” are not related to
any order, so in case the OYH command is
used, set 1 as a dummy.
13.1 Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT210 and is the answer not in the manual?

YOKOGAWA WT210 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT210
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals