<2.Specications>
2-5
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
STANDARD SPECIFICATIONS (High
Temperature Humidity Analyzer)
Sample Gas Temperature: 0 to 700°C (Probe
only)
It is recommended to mount the cell
using inconel cell-bolts when the
temperature is greater than 600°C.
Sample Gas Pressure: - 5 to + 20 kPa (When
the pressure in the furnace exceeds
3 kPa, it is recommended to use
pressure compensated type. When
the pressure in the furnace exceeds
5 kPa, pressure compensated type is
 

allowed.
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
Probe Material: SUS316
Ambient Temperature: -20 to +150°C
Reference Gas System: Natural Convection,
Instrument Air System (excluding
Natural Convection) It is recommended

by cooling to dew point -20°C or less,
and dust or oil mist are removed.

+approx. 50 kPA
Sample gas pressure + approx. 150 kPA (/CV:
with check valve
Wetted Material: SUS316, Zirconia, SUS304 or

(Inconel 600, 601)
Construction: Heater and thermocouple
replaceable construction. Non


IP66 (Achieved when the cable entry
is completely sealed with a cable
gland in the recirculation pressure
compensated version.)

Terminal Box Paint Color:

5.6BG3.3/2.9)
Finish: Polyurethane corrosion-resistance
coating
Gas Connection: Rc1/4 or 1/4 NPT (Female)
Wiring Connection: G1/2, Pg 13.5, M20 × 1.5,
1/2 NPT
Installation: Flange mounting
Probe Mounting Angle: Horizontal to vertically
downward. When the probe insertion
length is 2 m or less, installing at
angles from horizontal to vertically
downward is available.
When the probe insertion length
exceeds 2.5 m, mount vertically
downward (within ± 5°) and use a
probe protector.
Weight:
Insertion length of 0.4 m: approx. 6 kg (JIS 5K 65)
/ approx. 11 kg (ANSI 150 4)
Insertion length of 1.0 m: approx. 8 kg (JIS 5K 65)
/ approx. 13 kg (ANSI 150 4)
Insertion length of 1.5 m: approx. 10 kg (JIS 5K 65)
/ approx. 15 kg (ANSI 150 4)
Insertion length of 2.0 m: approx. 12 kg (JIS 5K 65)
/ approx. 17 kg (ANSI 150 4)
Insertion length of 3.0 m: approx. 15 kg (JIS 5K 65)
/ approx. 20 kg (ANSI 150 4)

Table of Contents

Other manuals for YOKOGAWA ZR802G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ZR802G and is the answer not in the manual?

YOKOGAWA ZR802G Specifications

General IconGeneral
BrandYOKOGAWA
ModelZR802G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals