<2.Specications>
2-15
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
Power Consumption: Max. 800 VA, approx. 330 VA for ordinary use.
Power supply 100V AC: Max. 160 VA (160 W), approx. 120 VA (approx. 100 W) for ordinary
use
Power supply 230 V AC: Max 550 VA (370 W), approx. 260 VA (approx. 100 W)
less (when a 1.25 mm
2
Construction: NEMA/CSA TYPE 4X (with conduit holes completely sealed with a cable gland)
Wiring Connection: eight holes
Installation: Panel, wall or 2-inch pipe mounting
Material:
Paint Color: Silver gray (Munsell 3.2PB7.4/1.2)
Finish: Polyurethane corrosion-resistance coating
Weight: Approx. 5 kg
Functions
Display Functions:
resistance tester.
junction temperature, maximum/minimum oxygen concentration, or the like
Indicates status such as warming-up, calibrating, or the like by the marks.
NE107 compliant 4-symbol display
Calibration Functions:
calibration direction on the touchscreen or contact, then it calibrates automatically
afterwards.
operation interactively with an LCD touchscreen.
2
0.01 vol% O
2
)
2
0.01 vol% O
2
)
Use N
2
-balanced mixed gas containing 0 to 10% scale of oxygen, and 80 to 100 % scale
of oxygen for standard zero gas and standard span gas respectively.
Purging Function: Before warming up the detector, feed the span gas for the set period of time to drain
condensed water out of the piping of calibration gas. Detector’s warming-up starts after the
set period of purging time elapses.
Blowback Function: To allow a periodic purging etc., open/close contact output in the set period of
Fault:
keeps turning on until the power shuts down.
error, data redundancy mismatch
Alarm: