<2.Specications>
2-16
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
alarm, EMF stabilization time-up alarm, Cold junction temperature alarm, Thermocouple voltage alarm,
Input current alarm, Battery low alarm, Input-pressure alarm, Cell resistance alarm
C: Function Check
Data Logging Function: Stores following data to SD card or visualizes on the instrument display. SD
cell resistance tester, cell condition, NE107 status, etc.) Maintenance report (setup value,
calibration value etc.) can be saved to SD cards in CSV format. The stored data can be
copied to other converter by outputting user-setting parameters to SD cards.
Sensor Self-diagnosis Function:
hour/minute) measurement interval (day/time)
Display and setting content:
Measuring Related Items: Oxygen concentration (vol% O
2
), output current value
Display Items: Cell temperature (°C), thermocouple reference junction temperature (°C), maximum/minimum/average
oxygen concentration (vol% O
2
grades), heater on-time rate (%), calibration record (twenty times), time (year/month/day, hour/minute)
Calibration Setting Items: Span gas concentration
(vol% O
2
), zero gas concentration (vol%
O
2
), calibration mode (automatic, semi-automatic, manual), calibration type
and method (zero-span calibration, zero
calibration only, span calibration only),
stabilization time (min. sec), calibration
time (min. sec), calibration interval (day/
hour), starting time (year/month/day,
hour/minute)
Output Related Items: Analog output/output mode
selection, output conditions when
warming-up/maintenance/calibrating
(during blowback)/abnormal, oxygen concentration at 4 mA/20 mA (vol%
O
2
), time constant.
Alarm Related Items: Oxygen concentration high
alarm/ high-high alarm limit values (vol%
O
2
), oxygen concentration low alarm/
low-low alarm limit values (vol% O
2
),
oxygen concentration alarm hysteresis
(vol% O
2
), oxygen concentration alarm
detection, alarm delay (seconds)
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 to 3 (Fault,
high-high alarm, high alarm, low alarm, low-low alarm, maintenance, calibrating, range
switching, warming-up, calibration gas pressure decrease, temperature high alarm,
temperature high alarm, pressure high alarm, pressure low alarm, test result from a cell
l Standard Specication (High Temperature Humidity Analyzer)
Display: LCD color display of size 320 by 240 dot with touchscreen.
Analog Output: