PRODUCT DESCRIPTION
Carl Zeiss Technical data Auto Focus equipment
18 425131-7044-001 06/2013
2.7.3 Operating data
Sensor module with IR-LED
Complete device .................................................................................................................... LED class 1
Internal LED .......................................................................... IRED 850 nm / max. 950 mW / LED class 3B
conforming to DIN EN 60825-1:2003
Auto Focus ....................................................................... Free group conforming to DIN EN 62471:2009
fc12 focus controller
Type of operation site ............................................................................................. Closed room facilities
Protection class .................................................................................................................................... I
Ingress protection rating .................................................................................................................. IP 20
Electrical safety .................................................................................... to DIN EN 61010-1 (IEC 61010-1)
conforming to CSA and UL regulations
Overvoltage category ............................................................................................................................ II
Radio interference suppression ................................................................................ to EN 55011 Class A
Noise immunity ..........................................................................................................to DIN EN 61326-1
Line input voltage ................................................................................. 100 V AC to 240 V AC (± 10 %)
No line voltage conversion required!
Line frequency........................................................................................................................ 50 / 60 Hz
Power consumption ..............................................................................................................max. 80 VA
Fuses according to IEC 60127
fc12 focus controller ..................................................................................... T 2.0 A/H; 250 V; 5x20 mm
Operating data for Axio Imager 2 / Axio Imager Vario, motorized with external VP232-2 power
supply pack.
Type of operating site ............................................................................................. Closed room facilities
Protection class .................................................................................................................................... I
Ingress protection rating .................................................................................................................. IP 20
Electrical safety .................................................................................... to DIN EN 61010-1 (IEC 61010-1)
conforming to CSA and UL directives
Overvoltage category ............................................................................................................................ II
Radio interference suppression ................................................................................. to EN 55011 Class B
Noise immunity ..........................................................................................................to DIN EN 61326-1
Line input voltage with external power supply VP232-2 ................................... 100 V AC to 240 V ±10 %
No line voltage conversion required!
Line frequency........................................................................................................................ 50 / 60 Hz
Max. power consumption Axio Imager, motorized ................................................................ max. 190 VA