9650-0309-01 Rev. E ZOLL AED Pro Service Manual 1–1
Chapter 1
Maintenance Tests
This chapter describes the procedures for routine testing of the AED Pro
®
unit.
A test typically consists of multiple steps, each of which requires a pass/fail rating. For each
step, perform the specified action and then verify that the expected results occur. If all expected
results occur, assign a “Pass” rating to the step; otherwise, assign it a “Fail” rating.
This chapter includes the following inspection and test procedures:
• “1.0 Physical Inspection” on page 1-3
• “2.0 Power Supply Test/Current” on page 1-4
• “3.0 Power ON Test” on page 1-5
• “4.0 Pediatric Mode/Electrode Connector Type Test” on page 1-6
• “5.0 Defibrillator Energy Test” on page 1-7
• “6.0 Battery Type Test” on page 1-9
• “7.0 Establish Communication” on page 1-10
• “8.0 Internal Discharge Test” on page 1-11
• “9.0 Patient Impedance Calibration” on page 1-13
• “10.0 ECG Lead-Off Detection” on page 1-15
• “11.0 ECG Offset Calibration” on page 1-16
• “12.0 Real Time Clock Test” on page 1-18
• “13.0 Write Calibration to Flash” on page 1-20
• “14.0 Audio Recording Test (if option installed)” on page 1-21
• “15.0 Read Error Log - Final” on page 1-22