Physical Inspection
9650-0309-01 Rev. E ZOLL AED Pro Service Manual 1–3
1.0 Physical Inspection
Use this procedure to ensure that the unit shows no signs of damage or excessive wear.
Tools Needed
None.
Test Setup
None.
Test Procedure
Observe the following: Pass Fail
1.1 Is the unit clean and undamaged?
1.2 Are all signs of wear reasonable and not excessive?
1.3 Are the front panel and housing undamaged and free of cracks?
1.4 Are input connectors clean and undamaged?
1.5
Are all cables free of cracks, cuts, exposed or broken wires? Are the pads
sealed within the packaging? If you are using an AED Pro 3-lead cable, is
it free of cuts or other?
1.6
Verify that the electrodes have not expired?
1.7
Is battery housing free of cracks and damage?