CHAPTER 1 MAINTENANCE TESTS
1–10 www.zoll.com 9650-0309-01 Rev. E
7.0 Establish Communication
Equipment Needed:
• Power Supply 12V 10A,
• Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
• AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF),
• Defibrillator Analyzer (e.g. QED-6 or Impulse 4000) or equivalent,
• PC with ZOLL RescueNet Code Review and ZAS (ZOLL Administration Software, P/N
7771-000012-01),
• ZOLL USB IrDA port (P/N 9355-0807).
Setup:
• Start the ZAS on the computer; make sure the connection is established before starting the
AED Pro unit.
• Note: For best transmission results, the ZOLL USB IrDA port (P/N 9355-0807) must be
facing the IrDA port of the unit with a clear 10 to 18-inch line-of-sight.
• Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Test Action Expected Result(s) Pass Fail
7.1
If the unit is on, press the On/Of
button for 1 second to turn off
the unit.
Unit powers off.
7.2
Press and hold the On/Off
button for at least 7 seconds,
then release it.
The AED Pro unit establishes
contact with the Computer
within 5 seconds and the unit
issues the voice and text prompt
message NON RESCUE
MODE, COMMUNICATION
ESTABLISHED.